When a picoampere is too coarse

A reading of 0.03 pA is hard to interpret at a glance, while 30 fA is plain. Electrometers, ion detectors, photomultiplier tubes and low-leakage semiconductor tests report currents in femtoamperes. Spec sheets for picoammeters list resolution and noise in fA, and a transistor leakage test might guarantee under 100 fA. This page multiplies a picoampere value by 1,000 so you can place a reading beside such a specification. It is useful for anyone moving between an instrument that displays pA and a data sheet quoted in fA, and it removes the usual slip of shifting the decimal point the wrong way.

Why the femtoampere figure is always the bigger number

Pico is 10^-12 and femto is 10^-15 in the SI Brochure (BIPM) and NIST SP 811, so 1 pA is exactly 1,000 fA. The formula is fA = pA × 1,000. Move the decimal point three places to the right: 0.03 pA is 30 fA, 2.5 pA is 2,500 fA and 0.4 pA is 400 fA. In the reverse direction, 750 fA is 0.75 pA and 12,000 fA is 12 pA. The result is always a larger number than the input, since the femtoampere is the smaller unit. If the page returns a smaller number, the menus are reversed. Floating-point display may show a trailing digit such as 9,999.999999999998 for the default; the intended value is 10,000 fA.

Ten picoamperes on the default screen

With 10 pA the page returns 10,000 fA, though a floating-point display may show a final digit slightly below that. Ten thousand femtoamperes is a size you meet in the input bias of a precision amplifier, though a data sheet would normally print it as 10 pA. The femtoampere form takes over once the number falls below one picoampere. A leakage of 0.5 pA is 500 fA, and 0.02 pA is 20 fA. Counting in fA is a signal that the current is under a picoampere, so the unit also tells the reader that special care was needed to measure it. When a table has entries from 100 pA down to 5 fA, one unit for the whole column is easier to sort.

Counting charges per second

The elementary charge is 1.602176634 × 10^-19 C, defined exactly in the SI. A current of 1 fA therefore transports about 6,241 electrons each second, and 1 pA transports about 6.24 million. A reading of 100 fA means roughly 624,000 electrons per second. At such levels the current is made of individual carriers, and the measurement shows shot noise, which grows with the square root of the current and the bandwidth. This explains why averaging time matters: a 10 fA signal needs seconds of integration to rise above noise. The conversion on the page keeps the numbers consistent, and the physics shows why the readings are so sensitive.

Guarding, triboelectric and dielectric effects

Femtoampere measurements fail without careful fixtures. The surface of a circuit board, a connector or a cable can leak more than the signal. Low-noise triaxial cables and guarded inputs hold the shield at the same potential as the signal conductor, so leakage across the insulation carries almost no voltage and almost no current. Bending a normal coaxial cable creates triboelectric charge that appears as a current of picoamperes. Dielectric absorption in an insulator makes readings drift for minutes after a voltage step. Wait for settling before you record a value, since the first reading after a voltage change is rarely the final one. The page cannot fix those effects, and a reading in fA is only as good as the setup behind it. Test engineers who characterize a component at 100 fA often spend more time on the fixture than on the measurement itself: cleaning the board with solvent, baking it to remove moisture, shielding the whole assembly in a metal box and leaving the setup powered for an hour before they take the first number. Each of those steps lowers the background that sits under the signal, and a log that lists them lets someone else reproduce the result. Skipping any one of them can add a few hundred femtoamperes of apparent current.

Photon counting and detector dark current

A photomultiplier tube or a photodiode produces a current equal to the number of detected photons per second multiplied by the charge per event and by the gain. A silicon photodiode with a dark current of 50 fA at 25 °C has a floor equal to 0.05 pA, and cooling can lower it. For 1 pW of light at 650 nm, a sensor with responsivity of 0.4 A/W gives 0.4 pA, which is 400 fA, well above that floor. These figures are illustrative. Comparing a measured current with the dark current in the same unit is the first step in judging whether a low-light signal is real or only background.

Semiconductor and sensor leakage specs

Chip makers specify gate leakage, input bias and off-state leakage for ultra-low-power parts at room temperature, and the values change sharply with heat. A typical rule is that leakage doubles for every 10 °C rise, so a part with 100 fA at 25 °C may show about 800 fA at 55 °C. Data sheets quote a typical and a maximum, and a measurement above the maximum signals a problem with the part or the setup. By converting both numbers to fA, you can see the ratio instantly. The result is only useful if the temperature and bias voltage match the conditions of the specification. For example, a leakage of 0.2 pA measured at 3.3 V on the bench says nothing about the same part at 1.8 V, because leakage depends strongly on the applied voltage. Record the supply, the temperature and the pin tested in the same row as the number, and the next person can repeat it. If a lot of parts are tested, keep every reading in femtoamperes in one column so that outliers stand out at once.

Noise and fixture leakage at femtoampere level

At femtoampere level a conversion is the easy part. The page does not measure current, subtract noise, correct for leakage in a fixture or judge whether a device passes a test. Rely on the instrument manual, the data sheet and a procedure that suits your setup. A conversion cannot add accuracy that the measurement lacks.