EMF meters read in milligauss

In the United States, handheld meters for household and workplace magnetic fields have long displayed milligauss, and many instrument manuals, inspection reports and older documents keep that unit. Scientific papers and international standards use teslas. A person reading both has to convert between them. This page multiplies the tesla value by 10,000,000 and returns milligauss, so the default 10 T becomes 100,000,000 mG. The two-step path is clear: 1 T is 10,000 G, and each gauss is 1,000 mG. The factor is exact, because the gauss is defined as 10^-4 T and milli means one thousandth. Since meters in this class often show µT on a second line, the page helps check that both displays agree.

Three ways to the same number

The conversion can be done in a few ways, and checking one against another catches slips. Multiply by 10,000 to reach gauss and then by 1,000 to reach milligauss. Or use microteslas as a bridge: 1 T is 1,000,000 µT, and 1 µT is 10 mG, so 1 T is 10,000,000 mG. A reading of 0.5 T is 5,000,000 mG by either path. The microtesla bridge is handy when a meter has a switchable display, because many meters show µT and mG on the same screen, and the ratio of 10 between them is easy to remember. When both paths agree, the arithmetic is almost certainly right.

A hundred million milligauss

With 10 T entered, the page returns 100,000,000 mG, which is 1 × 10^8. Smaller inputs work the same way: 0.001 T is 10,000 mG, and 0.0000025 T is 25 mG. If your answer is 100,000, you stopped at gauss and skipped the last prefix step. The numbers at this level are far from anything an EMF meter handles, since such meters are built for the weak fields near wiring and appliances. The default shows the size of the factor, since a 10 T field is eight orders of magnitude above one milligauss.

Earth's field in milligauss

The Earth's field at the surface is about 0.25 to 0.65 G, which is 250 to 650 mG, or 25 to 65 µT. That is a good reference when you use a milligauss meter that responds to static fields. A reading near 500 mG on such an instrument outdoors is normal, while a reading many times higher indicates a magnet or steel object near the probe. A meter that measures only alternating fields will read near zero for the Earth's field, because the Earth's field does not vary with time. Know which type of meter you hold before interpreting a number. A magnetized tool, a steel chair leg or a speaker magnet near the probe can shift a static reading by far more than a few mG.

Home and workplace surveys

An inspector surveying a room with an alternating-field meter walks along walls, near the electrical panel, close to appliances and over wiring, and records readings in mG. The values vary with the load on the circuits, so a reading taken at one moment may differ from another, and they fall off quickly with distance from the source. Reports should note the position, the distance from the source and the time. This page does not interpret those numbers. It helps when a reading in milligauss has to be compared with a value in microtesla or tesla from another document, for example a table in a guideline or a research paper. Photographing the meter display next to the source is a useful habit, since it fixes the value and the unit together. Note the room, the distance and the time of day beside each number.

Single-axis and tri-axis meters

A single-axis meter reads only the component along its sensing axis, so rotating it changes the reading, and the true maximum is found by turning it. A tri-axis meter measures three perpendicular components and reports the magnitude, which is the square root of the sum of the squares. A tri-axis result is therefore at least as large as any single-axis reading at the same point. When two people measure the same spot with different meters, a difference in the numbers may come from the axes and not from the unit. The conversion to or from teslas applies to whichever value the meter displays. Orientation also matters on a single-axis meter, which is why careful surveys turn the probe in three directions.

mG and mT are far apart

The symbols mG and mT differ by one letter and by a factor of 10,000. One millitesla is 10 G, which is 10,000 mG. A value of 3 mT is 30,000 mG, and 3 mG is only 0.0003 mT. Reading a field meter display set to mT as if it were mG would overstate the field by four orders of magnitude, and the opposite slip would understate it. A habit that prevents the error is to write the symbol with every number you record, and to convert at the end with a tool like this one. Check the unit on the screen, since many meters can be switched between them. A converted check helps: 3 mT must be 30,000 mG, and 3 mG must be 0.3 µT.

From milligauss back to teslas

Choose milligauss as the source and teslas as the target to divide by 10,000,000. A reading of 2.5 mG is 2.5 × 10^-7 T, which is 0.25 µT. A reading of 500 mG is 0.00005 T, which is 50 µT, a typical Earth-field figure. A reading of 5 mG, by comparison, is 0.5 µT, one hundredth of that, so surveys of weak fields call for a meter with fine resolution. For documents, microteslas are a sensible middle ground, since 1 mG is 0.1 µT and the numbers stay readable. State the original reading and the converted one together, such as 500 mG (50 µT), so a reader who uses either unit can follow. The microtesla figure is also the one most international documents use for weak fields.

Not an exposure assessment

The page changes a unit and does not judge exposure. Guidelines for magnetic field exposure depend on frequency, duration and population, and they are published by national and international bodies that update them. If you need to assess a situation, consult the current guideline and a qualified professional, and report the measurement conditions with the number. Field values near wiring also change with the load, so a single reading is a snapshot.