Where teraohm values are used
A teraohm is a million megaohms, a resistance so high that only specialized instruments measure it. Teraohm figures appear in the specifications of electrometers, in data sheets for guard-ringed test fixtures, and in tests of insulating materials such as PTFE and polyethylene, where volume resistivity values are enormous. High-value resistors of 100 GΩ to 1 TΩ are used as feedback elements in picoammeters and in some radiation and photodiode amplifiers. A report that also lists ordinary parts in kΩ will have two columns that differ by nine orders of magnitude. This page converts a kΩ value to TΩ so that both sit on one scale.
One billion kilohms per teraohm
Tera is 10 to the twelfth and kilo is 10 to the third, so 1 TΩ equals 1,000,000,000 kΩ. The default 10 kΩ becomes 0.00000001 TΩ, or 1 times 10 to the minus eighth. A value of 100,000 kΩ (100 MΩ) is 0.0001 TΩ, and 50,000,000 kΩ (50 GΩ) is 0.05 TΩ. A value of 1,000,000,000 kΩ is exactly 1 TΩ. Going the other way, 2 TΩ is 2,000,000,000 kΩ. Because the number of zeros is large, keep the exponent in view: kΩ values near 10 to the ninth correspond to TΩ values near 1.
Leakage current at very high resistance
Ohm's law shows what a teraohm means in practice. With 100 V across 1 TΩ, the current is 100 pA, one ten-billionth of an ampere. With 10 V across 5 TΩ, the current is 2 pA. These currents are small enough that a nearby board contaminated with flux residue, or a cable with poor insulation, can swamp them. Measurements at this level need shielded cables, guarded connections and dry air. The converter changes units. The measurement skill is in the setup, and the literature from electrometer makers describes the methods in detail. Even the cable matters: a few centimeters of dirty insulation between two terminals can carry more current than the part under test, so guarded fixtures, clean connectors and sufficient settling time after the voltage is applied are part of the procedure.
Guards, shields and settling time
At teraohm levels the measurement fixture is as important as the instrument. A guard ring is a conductor held at the same potential as the measured terminal, placed so that surface leakage flows into the guard instead of into the meter. A metal shield around the sample blocks stray electric fields that would induce currents of the same size as the signal. Settling is slow too: a 1 TΩ resistor with 10 pF of cable capacitance has a time constant of 10 seconds, and a reading needs several time constants to stabilize. A 100 GΩ source with 100 pF has the same 10 seconds. Waiting for a steady value, and noting how long you waited, makes a teraohm reading repeatable.
Feedback resistors in picoammeters
A transimpedance amplifier turns a tiny current into a voltage using a feedback resistor. A 1 GΩ feedback resistor gives 1 V per nanoampere, and a 100 GΩ resistor gives 100 mV per picoampere. Pushing toward TΩ values raises the gain further, but the noise from the resistor, its voltage coefficient and its leakage over the board surface all limit what is possible. Designers pick the resistance that produces a convenient output for the expected current, and often use a second, smaller resistor for ranges with larger currents. Comparing a 1 TΩ feedback value with a 10 kΩ input resistor, 1,000,000,000 kΩ against 10 kΩ, a ratio of 100,000,000 to 1, shows how much gain the stage can offer.
Insulating materials and surface resistance
Plastic and ceramic insulators are specified by resistivity, in ohm-meters or ohm-centimeters, not by the resistance of a single part. A test cell converts the reading into a resistivity using the electrode geometry. A raw reading of 3 TΩ taken on a sample is the resistance between the electrodes, 3,000,000,000 kΩ, and it depends on the thickness and area of the sample. Surface and volume resistance are distinct measurements, because current can travel across the surface or through the bulk. Standards describe the electrode arrangement and test voltage for each. Humidity can lower a surface reading by orders of magnitude, so a stated result needs its conditions. A sample that reads 3 TΩ after drying in a desiccator may read a few hundred GΩ after a day in humid air, which is a drop from 3,000,000,000 kΩ to a few hundred million kΩ. Record temperature, relative humidity, applied voltage and electrification time beside each reading, and keep the unit of every entry the same through the dataset.
What an insulation tester shows
Some insulation testers display values into the teraohm range, with limits that depend on the model and the test voltage. A reading of 2.5 TΩ at 1,000 V corresponds to a leakage current of 0.4 nA. Many testers also show a polarization index, the ratio of the reading at ten minutes to the reading at one minute, which indicates how the insulation behaves as it charges. The index is a ratio and has no unit. The converter handles only the unit for the resistance figure. Acceptance values and test procedures belong to the standard and to the equipment manufacturer.
Notation in spreadsheets
Resistance columns that range from 10 Ω to 5 TΩ cannot be read as plain numbers. Many labs write values in engineering notation, for example 4.7E3 for 4.7 kΩ and 2.5E12 for 2.5 TΩ, and keep the unit in the column header. This page returns the TΩ value in decimal form, so a result such as 0.00000001 appears for the default. If your sheet shows it as 1E-08, the value is the same. Sort and filter by the number rather than by text, and store raw readings beside converted ones for later checking.
Very high resistance drifts with voltage
The conversion restates the same known resistance in a compatible unit. It does not calculate installation limits, equipment ratings, protection, material response or compliance, and it does not tell you whether a reading is acceptable. Very high resistances change with voltage, temperature, humidity and time under test, so two readings in the same unit can differ for legitimate reasons. Read the applicable standard and the instrument manual before drawing conclusions from any value.