Noise specs and signal levels
The noise of an amplifier is quoted in nanovolts, usually as a density in nanovolts per root hertz, while the signal it amplifies may be a few millivolts. A designer who picks an op amp for a sensor interface, a technician who reads a nanovoltmeter, or a student who compares a lock-in measurement with its noise floor needs both on one scale. This page multiplies the millivolt value by 1,000,000 and returns nanovolts. The six decades between the two units match the gap between a signal and the fluctuations that limit how well it can be measured, so the converted number tells you immediately whether the signal is large or small compared with the noise.
Six powers of ten
Milli is 10^-3 and nano is 10^-9, so a millivolt holds 10^6 nanovolts. The volt is one joule per coulomb in the BIPM SI Brochure, and the prefixes are defined, so the conversion has no uncertainty. For a check, convert in two steps through microvolts: multiply by 1,000, then by 1,000 again. A reading of 1.5 mV becomes 1,500 µV, then 1,500,000 nV. Because the result is a long number, the digit count is a quick test: a one-digit millivolt value should give seven digits in nanovolts. Fewer digits suggest that the wrong unit was selected for one of the two fields.
Ten millivolts in nanovolts
The default 10 mV gives 10,000,000 nV. A smaller value of 0.002 mV equals 2,000 nV, and 1.5 mV equals 1,500,000 nV. In the reverse direction 3,500 nV equals 0.0035 mV and 20 nV equals 0.00002 mV. Anchors help here: 1 µV is 1,000 nV, and 1 nV is a millionth of a millivolt. A long result is easier to read in groups of three, or as an exponent such as 1 × 10^7 nV. Keep the original millivolt entry next to the converted value in a log, so that a typo in the long number can be caught by comparing the two. For instance, a lock-in amplifier sensitivity setting of 10 mV full scale is 10,000,000 nV, and an output resolution of 0.01 percent of that range is 1,000 nV, which tells you the smallest step the instrument can show on that setting.
Reading an op-amp noise density
Low-noise operational amplifiers list voltage noise density in the range of 1 to 10 nV per root hertz, for example 2 nV/√Hz. The total noise over a bandwidth B is the density times the square root of B. For 2 nV/√Hz over a bandwidth of 20,000 Hz the result is 2 × 141.4 = 282.8 nV rms, which is 0.0002828 mV. Compared with a 1 mV sensor signal, which is 1,000,000 nV, that noise is about 0.03 percent, a comfortable margin. With a 10 µV signal the same noise would be about 2.8 percent. The exercise is illustrative, and real parts add current noise, 1/f noise and the thermal noise of the source resistance.
Nanovoltmeters and thermal EMFs
A nanovoltmeter resolves signals down to a nanovolt or less, and its use is common in measurement of resistance of low-resistance samples, in thermoelectric tests and in metrology labs. At that level, unwanted thermal voltages from the junctions of dissimilar metals, often a microvolt per degree or more, can exceed the signal. A reading of 1,500 nV is 0.0015 mV, which is far below the 0.1 mV resolution of a handheld multimeter. Wiring technique matters: use copper-to-copper connections, keep temperature stable and shield from air drafts. The conversion lets you check that a measurement system is built for the scale of the signal before anything is connected.
Converting nanovolts back to millivolts
To convert in the other direction, choose nanovolts as the source and millivolts as the target. A value of 250,000 nV equals 0.25 mV and 3,500 nV equals 0.0035 mV. The page divides by 1,000,000. For a spreadsheet of noise measurements it may be easier to move to microvolts, where 3,500 nV is 3.5 µV. The same page offers microvolts and picovolts, so a chain of conversions can step to the unit with the most readable digits. Many bench tools display automatically in the best prefix, and a log that has both nV and µV entries should be normalized to one unit before it is averaged or plotted. Averaging 3,500 nV and 2.1 µV without converting gives a meaningless number, while converting both to nanovolts gives 3,500 and 2,100, and an average of 2,800 nV. Normalizing the units first is the safest habit in any logging script.
Prefix traps at the small end
Nanovolts and microvolts are easy to swap in a quick read, and the difference is a factor of 1,000. Another trap is the density unit nV/√Hz, which is not a voltage and must not be converted as one. It becomes a voltage only after multiplication by the square root of the bandwidth. A third trap is peak versus RMS, since noise values are almost always RMS while a peak-to-peak figure is about six times larger for Gaussian noise. A third trap is peak versus RMS, since noise values are almost always RMS while a peak-to-peak figure is about six times larger for Gaussian noise. A noise of 282.8 nV RMS would then show about 1,700 nV peak to peak on a scope, which is 0.0017 mV, and a reader who compares the wrong pair of numbers misjudges the margin by a large factor. When a datasheet states noise in µV peak to peak over 0.1 to 10 Hz, treat it as a different quantity from a density, and convert only the prefix.
Writing a 2 nV noise figure without false digits
The conversion is exact, so the number of significant digits follows the input. A noise figure given as 2 nV has one significant digit, and a result of 0.000002 mV should not be written with more. The page does not compute noise bandwidth, resistor thermal noise, amplifier gain or measurement uncertainty. It also does not know whether your number is a mean, an RMS value or a peak, so label it before converting. It restates one voltage in another unit. For design decisions, use the datasheets, the instrument specifications and a proper noise analysis. A spreadsheet that holds the signal in millivolts and the noise terms in nanovolts, with one conversion column between them, is usually enough to see where the largest contribution comes from.