Water chemists read conductivity in µS/cm. Materials scientists, battery researchers and electrochemists report conductivity in S/cm, and metals appear in S/m or MS/m. A reading from a bench conductivity meter has to be converted when it goes into a paper or a data sheet that uses S/cm. This page converts µS/cm to S/cm, a factor of one million. It also helps in reverse, when a literature value in S/cm must be compared with the limited range of a field meter. The explanation below covers the factor, resistivity, the cell constant and a few reference magnitudes that make the numbers meaningful.

Six decimal places of difference

One siemens is 10^6 µS, so one µS/cm is 10^-6 S/cm. The page computes S/cm = µS/cm × 0.000001. The default 10 µS/cm gives 0.00001 S/cm. A reading of 500 µS/cm gives 0.0005 S/cm, and a reading of 0.5 µS/cm gives 5 × 10^-7 S/cm. The length unit is the same on both sides, so only the prefix changes. Both quantities carry the unit of siemens, so the meter display and the lab sheet describe the same property. A direction check helps, because the number in S/cm is a million times smaller than the number in µS/cm.

Electrolytes in batteries

Researchers describe liquid electrolytes for lithium-ion cells by ionic conductivity, and values on the order of 10 mS/cm are typical for common organic carbonate solutions at room temperature. That is 0.01 S/cm, or 10,000 µS/cm. A meter on a bench with a cell designed for such solutions reports the value in mS/cm, and the paper wants S/cm. The conversion is a single step. Solid electrolytes under study often have conductivities far lower, and values like 10^-3 S/cm or 10^-4 S/cm are discussed as targets. Seeing 0.001 S/cm as 1,000 µS/cm gives a feel for how those values compare with ordinary tap water at a few hundred µS/cm. The comparison also shows why a water meter is the wrong instrument for such materials, since its upper range stops far below the values of interest and its electrodes are not designed for organic solvents.

Resistivity as the reciprocal

Conductivity and resistivity are reciprocals. A conductivity of 0.01 S/cm corresponds to a resistivity of 100 Ω·cm. Water purification systems often display resistivity in MΩ·cm, and 18.2 MΩ·cm, the value for ultrapure water at 25 °C, corresponds to about 0.055 µS/cm, which is 5.5 × 10^-8 S/cm. Converting between the two requires care with prefixes. A tool that shows resistivity can be read against this page by taking the reciprocal first, in Ω·cm, and then entering the result as S/cm, or by first converting to µS/cm and then using the factor of one million.

Metals and the copper reference

Metals have conductivities that are huge in comparison with solutions. The International Annealed Copper Standard sets 100 percent IACS at a resistivity of 1.7241 µΩ·cm, which is a conductivity of 58.0 MS/m. In S/cm that is about 5.8 × 10^5 S/cm, and in µS/cm it is 5.8 × 10^11. Tables of materials often list metals in MS/m, and chemists will meet S/cm. The page lets you place a solution at 500 µS/cm, or 0.0005 S/cm, beside copper to see that the ratio is on the order of 10^9. Entering 5.8 × 10^5 S/cm with the reverse units gives the corresponding number in µS/cm.

The cell constant in S/cm

A conductivity cell measures conductance G in siemens, and the conductivity equals G times the cell constant K, where K is the electrode spacing divided by the electrode area, in cm^-1. A cell with K = 1.0 cm^-1 that reads 500 µS gives 500 µS/cm, or 0.0005 S/cm. A cell with K = 0.1 cm^-1 reads ten times more conductance for the same solution, which suits very pure water, while a cell with K = 10 cm^-1 suits strong electrolytes. Meters apply K automatically, and the unit appears on the display. The conversion on this page is applied after that step, to the finished conductivity value.

Reading values in scientific notation

Small S/cm values are usually written in scientific notation, such as 5 × 10^-7 S/cm. Type them with an exponent if the field allows it, or in full decimal form. The default result, 0.00001, is 1 × 10^-5 S/cm. A value with a negative exponent of six or more relative to a µS/cm reading belongs here. If the exponent in the result differs from the one you expect by three, the µS/cm value probably came from a meter in mS/cm. Conductivity values in papers also carry a temperature, usually 25 °C, which the conversion leaves unchanged. Record it next to the number. Many papers also give the electrode spacing, the frequency of the measuring signal and the sample composition, since ionic conductivity measured with alternating current can differ from a direct-current value for the same sample. Copying those details along with the converted figure keeps the number useful to the next reader.

From S/cm to µS/cm

Select siemens per centimeter as the source and microsiemens per centimeter as the target to multiply by 1,000,000. A value of 0.01 S/cm becomes 10,000 µS/cm, and 0.0005 S/cm becomes 500 µS/cm. Metals work the same way, although the results are enormous, such as 5.8 × 10^11 µS/cm for copper. Because the page prints large numbers in scientific notation, copy them with the exponent. When you transcribe into a spreadsheet, store the number and the unit in separate columns, so that a later reader does not confuse S/cm with S/m, which differ by a factor of 100. A conductivity of 0.01 S/cm is 1 S/m, and a value of 5.8 × 10^5 S/cm is 5.8 × 10^7 S/m, the figure many materials tables print for copper. Checking one such pair against a table is a good way to confirm that your units line up.

Material data sheets and measurement conditions

The page only restates a conductivity in S/cm. Whether a material suits a job is settled by its data sheet and the test method behind the figure. It does not model how conductivity changes with temperature, concentration or frequency, and it does not decide whether a material suits an application. Published conductivity values depend on the measurement method, temperature and sample preparation, so quote the source with each figure. For electrical safety questions, rely on qualified professionals and the standards that apply to the installation, not on a converted number.