Electrical conductivity: units and measurement guide

Electrical conductivity is a material quantity, distinct from an object’s conductance.

Measurement basis

Temperature and composition can affect a measured conductivity. Rescaling its unit does not establish those material conditions.

Definition conventions

Temperature compensation is not performed. The conversion does not determine TDS, salinity or concentration.

Conversion method

Account for both the conductance prefix and length denominator in the stored factor ratio.

Choose the conversion direction

Keep the full source unit and measurement conditions, then select the required conductivity unit.

Siemens per meter as the SI unit

Electrical conductivity is a material property that tells how well a material carries electric current. It is the reciprocal of resistivity, σ = 1/ρ, and the SI unit is the siemens per meter (S/m). A material with a conductivity of 1 S/m allows a current of 1 A through a cube one meter on a side under a potential difference of 1 V across opposite faces. Because geometry is built into the unit, conductivity can be compared across samples of any shape, which is the reason chemists, hydrologists and electrical engineers all use it. The related unit for objects, the siemens, describes one specific piece of wire or one specific cell, while S/m describes the material itself.

Why water analysts use microsiemens per centimeter

In water testing, siemens per centimeter and its submultiples are customary. The conversion between the two length bases is a factor of 100, since there are 100 cm in a meter: 1 S/m equals 0.01 S/cm, which is 10 mS/cm, and 1 mS/cm equals 0.1 S/m. For the microsiemens, 1 µS/cm equals 100 µS/m, or 0.1 mS/m. A groundwater sample at 500 µS/cm is therefore 50 mS/m, or 0.05 S/m. Mixing S/m and S/cm labels on instruments and reports is a common cause of factor-of-100 errors. Writing out both forms of the same value, such as 1,413 µS/cm equals 141.3 mS/m, helps catch the slip.

Typical readings from pure water to seawater

Ultrapure water has a conductivity near 0.055 µS/cm at 25 °C, corresponding to a resistivity of about 18.2 MΩ·cm, a figure used in semiconductor and laboratory water specifications. Distilled water often reads from 1 to 5 µS/cm because of dissolved carbon dioxide. Tap water commonly falls between about 50 and 800 µS/cm depending on the source, while seawater is around 50 mS/cm, or about 5 S/m. Brines and process solutions can reach hundreds of mS/cm. A calibration standard widely used for probes is 0.01 mol/L potassium chloride, which has a conductivity of about 1,413 µS/cm at 25 °C. These landmarks help judge whether a converted number lands in a believable range.

Temperature compensation

Conductivity of an electrolyte rises with temperature, by roughly 2 percent per degree Celsius for many natural waters. For that reason, instruments report readings referred to a standard temperature, usually 25 °C, and the converted value is meaningless unless the temperature reference is stated. A raw reading of 600 µS/cm at 15 °C corresponds to roughly 750 µS/cm at 25 °C using the 2 percent rule, an illustrative figure that actual compensation tables refine. Seawater practice uses a more elaborate practical salinity scale. Converting units does not change the temperature, so a document should always keep the reference temperature with the value. Metals behave in the opposite way: their conductivity falls as temperature rises.

Conductivity of metals

Metals have conductivities millions of times higher than water, and the numbers are given in megasiemens per meter. Annealed copper at 20 °C has a conductivity of 58.0 MS/m by the International Annealed Copper Standard, equal to a resistivity of 1.7241 × 10⁻⁸ Ω·m, and a material rated at 100 percent IACS meets that benchmark. Pure silver is a little higher, about 63 MS/m, and aluminum is around 35 to 38 MS/m, about 61 percent of copper. The percent IACS scale is used in the wire and cable industry and does not carry units. A conversion from 58 MS/m to S/m gives 58,000,000 S/m and to S/cm gives 580,000 S/cm, which shows how unwieldy the larger units become.

Relation to resistivity and total dissolved solids

Resistivity in ohm-meters is the reciprocal of conductivity in S/m, so 0.05 S/m is 20 Ω·m and 5 S/m is 0.2 Ω·m. Ohm-centimeters pair with S/cm, which explains the 18.2 MΩ·cm figure above. Meters sometimes also display total dissolved solids, TDS, estimated by multiplying conductivity by a factor commonly between 0.5 and 0.7 depending on the ion mix. That factor is an assumption, not a measurement, and the instrument setting should be recorded alongside the result. A conversion between conductivity units is exact arithmetic, whereas the step to TDS involves an assumption about composition, so the two should never be treated as equivalent operations.

Soil, geophysics and agriculture

Bulk electrical conductivity of soil is used to map salinity, moisture and clay content, and agricultural sensors report it in dS/m, decisiemens per meter, which is numerically identical to mS/cm. A soil extract at 4 dS/m is therefore 4 mS/cm, 4,000 µS/cm or 0.4 S/m, a level often used as a rough threshold when discussing saline soils. Geophysical surveys of rock and groundwater use mS/m, and values for dry rock can be very low, near 0.1 mS/m. Because fields differ in the unit they favor, a table that lists S/m, mS/cm and µS/cm side by side is more useful than any single factor. Always check whether the original is a bulk measurement of soil or a measurement of an extracted liquid.

Related quantities and sources

Conductivity should not be mixed up with conductance in siemens, or with thermal conductivity in W/(m·K), which describes heat transport with the same word. Molar conductivity, used for electrolyte solutions, has units of S·m²/mol and requires concentration. The BIPM SI Brochure defines the siemens, and NIST SP 811 lists S/m. For water testing, standard methods from ASTM and ISO describe cell constants, calibration and the 25 °C reference temperature. This page is concerned only with expressing the same conductivity in another unit, and it makes no statement about whether a particular reading is acceptable for drinking, irrigation or industrial use. Those limits come from the relevant regulator or process specification.

Cell constants and why probes read conductance first

A conductivity probe does not measure conductivity directly. It measures the conductance between its electrodes, and the instrument multiplies by the cell constant K, the ratio of electrode spacing to electrode area, with units of 1/cm or 1/m. A probe with K = 1.0 cm⁻¹ reading 1.413 mS gives 1.413 mS/cm, while a probe with K = 0.1 cm⁻¹ reading 0.1413 mS gives the same conductivity. Low-conductivity water, such as boiler feed water, is measured with K near 0.01 cm⁻¹, and brines with K around 10 cm⁻¹, so the electrodes stay within a workable range. A wrong cell constant setting produces a result off by exactly that factor, so confirm it before converting any unit.

Precision and checking

  • Retain the original measurement precision and conditions.
  • Keep the length denominator along with the siemens prefix.

Reference numbers are displayed with up to twelve significant digits. That display does not add precision to a rounded measurement. Keep source and target quantities compatible, confirm symbol case and regional conventions, and round the final result to precision justified by the input. Factors involving π or repeating fractions use finite numerical precision.

Choose a unit definition

Use a working converter

Source and verification scope

The reference values and examples use the versioned unit definitions shared with the linked converters. Source dates identify the stated access or verification scope. A reverse calculation checks arithmetic and direction, but it does not independently verify a definition. Computandum is independent of the cited organizations, and the citations do not imply endorsement of this guide or a real application.