Vacuum gauges and sensors speak different dialects

Vacuum equipment sold in Europe is usually rated in millibar. Pump data sheets, gauge ranges and process recipes carry numbers such as 10^-3 mbar. Sensors and research instruments that follow the SI strictly report in pascals, and the very small values end up in millipascals. Someone who sets up a coating chamber, a vacuum oven or a leak test often has both kinds of documents on the desk and has to line them up. This page divides the millipascal value by 100,000 and gives the millibar figure. It serves technicians, lab students reading a pump manual, and engineers checking whether a sensor range covers the pressure a process needs. Dividing by 100,000 is the whole job, so a gauge range in mbar lines up with a mPa figure directly.

Why the factor is one hundred thousand

A millibar is one thousandth of a bar, and the bar is defined as exactly 100,000 Pa, so 1 mbar is 100 Pa. A millipascal is one thousandth of a pascal, so 1 mbar equals 100,000 mPa. Going the other way, one millipascal is 0.00001 mbar. Both prefixes are milli, but they sit on different base units, which is why the numbers do not match one to one. The bar is not an SI unit, though the BIPM lists it as accepted for use with the SI, and NIST SP 811 gives the same relation. The bar is defined as 100,000 Pa, so no measured constant enters the result. A useful memory aid is that 1 Pa is 0.01 mbar, so every pascal is a hundredth of a millibar.

Ten millipascals as millibar

The default input of 10 mPa gives 0.0001 mbar, which is 10^-4 mbar. In pascals that is 0.01 Pa. Other values follow the same rule: 500 mPa is 0.005 mbar, and 100,000 mPa is 1 mbar. The output is far smaller than the input because the millibar is a unit 100,000 times larger than the millipascal. If the result looks bigger than the input, the from and to menus are swapped. Because many vacuum readings are written as powers of ten, it helps to convert the result into exponent form: 0.0001 mbar is 1 × 10^-4 mbar. Writing it that way makes it easy to compare with pump specifications and avoids losing a zero when the number is copied.

Where the number falls on the vacuum scale

Vacuum technology splits the pressure range into bands. A common division, used in ISO 3529, calls the range from atmospheric pressure to 1 mbar low vacuum, 1 mbar to 10^-3 mbar medium vacuum, 10^-3 to 10^-7 mbar high vacuum, and anything below 10^-7 mbar ultra-high vacuum. The default of 10^-4 mbar is therefore high vacuum, and a value of 0.1 mPa, which is 10^-6 mbar, is still inside that band. Typical mechanical pumps reach the medium range, while turbomolecular pumps are needed for the high range. This explains why datasheets for a given pump quote an ultimate pressure in millibar: it tells you at once which band the machine can serve.

A pump rated in millibar against a sensor in pascals

Suppose a rotary vane pump has an illustrative ultimate pressure of 5 × 10^-3 mbar, and the sensor on the chamber reports in millipascals. The pump limit is 0.5 Pa, which is 500 mPa. A sensor reading of 800 mPa, which is 0.008 mbar, tells you the chamber has not yet reached the pump limit, so a leak or outgassing is the likely cause of the gap. Another example is a freeze dryer or vacuum oven that runs near 0.1 mbar. That is 10 Pa, or 10,000 mPa. The numbers are illustrative arithmetic. The actual limits of your pump and gauge are printed in their manuals and depend on oil, seals, gas type and temperature.

From atmospheric pressure down to millipascals

Standard atmospheric pressure is 101,325 Pa, which is 1,013.25 mbar and 101,325,000 mPa. A weather map rounds this to about 1,013 mbar, or hPa, since one millibar equals one hectopascal. Weather pressure and vacuum pressure sit at opposite ends of an enormous range: the default 10 mPa is about ten billion times smaller than one atmosphere. That scale explains why gauges are built in stages, with a capacitance manometer at the upper end, a Pirani gauge in the middle and an ion gauge at the bottom, each covering a few decades. When a recipe quotes a number in millibar, check that your sensor is rated for that range before you trust the conversion.

Reading millibar back into millipascals

To change direction, choose millibar as the source and millipascals as the target. The page then multiplies by 100,000. A value of 1 × 10^-6 mbar is 0.1 mPa, and 0.02 mbar is 2,000 mPa. Vacuum figures are usually quoted with one or two significant digits, since gauge accuracy at high vacuum is often only about plus or minus a few percent, and sometimes worse for ion gauges. Keep the same number of digits in the converted result. If the chamber gas is not air or nitrogen, remember that many gauges read differently for other gases, so the number on the display may need a gas correction from the maker before it is converted.

Mixed units in a pump manual

Vacuum manuals are not always consistent. A pump may list its ultimate pressure in millibar on the front page, a flow curve in pascals on page twelve and a gauge range in Torr on the accessory sheet. Reading them side by side means converting each to one unit before comparing. A practical approach is to choose pascals as the working unit, since the SI relation is simple, and to convert the others with the defined factors: 1 mbar is 100 Pa, and 1 Torr is about 133.32 Pa. The millipascal figures from a research gauge fit into the same table. After the table is built, a pump limit of 0.005 mbar and a sensor limit of 100 mPa can be compared directly, and it becomes clear that the sensor covers the range the pump can reach.

What this conversion does not decide

The page converts the unit of a pressure value. It does not tell you whether a vacuum is good enough, correct a gauge for gas type, estimate pump-down time or check safety limits of a chamber. Absolute and gauge readings are different things, and vacuum work is nearly always absolute, so keep that label with the number. For equipment limits, read the manufacturer documents, and for safety, follow your own site procedures. Use the page only to put two pressure figures on the same scale.